Analysis of Patient Radiation Dose and Quality of Radiograph for Adult Thorax Diagnosis Use High Kilo Volt Method

Authors

  • Liberti Tarigan  Department of Physics, Universitas Sumatera Utara, Padang Bulan, Medan and Health Facility Security Center, Jl. KH Wahid Hasyim No. 15 Medan
  • Timbangen Sembiring Marhaposan S  Department of Physics, Universitas Sumatera Utara, Padang Bulan, Medan
  • Juliana   Department of Physics, Universitas Sumatera Utara, Padang Bulan, Medan and Health Facility Security Center, Jl. KH Wahid Hasyim No. 15 Medan
  • Hotromasari   Department of Physics, Universitas Sumatera Utara, Padang Bulan, Medan and Universitas Sari Mutiara, Jl. Captain Muslim No. 79 Medan
  • Berkat  Department of Elektromedic Engineering, Yayasan Binalita Sudama, Jl. Gdg PBSI, No. 1 Medan

Keywords:

high kV, adult thorax, contrast, incident air kerma.

Abstract

Quality of radiographic images of thorax is influenced by radiation quality. Hence research is needed related to the effect of high kV method for adult thorax diagnosis. The effect of the radiation quality on the image can be known through the density of the film resulted from the stepping of the stepwedge object. The effect of high kV use on doses was done by analyzing incident air kerma on two radiography planes with Philips and Toshiba brands. From this study it was found that using of high kV method will caused image contrast to decrease when compared with standard kV method (generally used). But the image of the lung bronchus will be seen more than the use of standard kV method. In the use of high kV method can be done with a low burden tube (mAs), Thorax consists of lung organs and heart. It is moves or hard to resist so it is suitable for use because it takes a short time. The use of high kV method decreases the patient's acceptable dose by still producing an optimal diagnosis of adult radiographs.

References

  1. Pearce, Evelyn C, 1992. Anatomy and Physiology For Paramedics, PT Gramedia Pustaka Utama.
  2. Jhon Ball, Tong Price, 1989. Chesney's Radiographic Imaging. London: Black well scientic.
  3. Bushong, Stewart C, St. Lois, Missouri, 2008. Radiologic Science For Radiologist.
  4. Meridith, WJ, Massey JB, 1977, Fundamental Physics of Radiology, John Wrigh & Son, Bristol

Downloads

Published

2018-07-30

Issue

Section

Research Articles

How to Cite

[1]
Liberti Tarigan, Timbangen Sembiring Marhaposan S, Juliana , Hotromasari , Berkat, " Analysis of Patient Radiation Dose and Quality of Radiograph for Adult Thorax Diagnosis Use High Kilo Volt Method, International Journal of Scientific Research in Science, Engineering and Technology(IJSRSET), Print ISSN : 2395-1990, Online ISSN : 2394-4099, Volume 4, Issue 9, pp.117-120, July-August-2018.