Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography Examinations in Two Hospitals

Authors

  • Daniel Ackom  Department of Physics, University of Cape Coast, Cape Coast, Ghana
  • Edem Sosu  Department of Medical Physics, School of Nuclear and Allied Sciences, University of Ghana, Atomic Campus, Accra, Ghana
  • Stephen Inkoom  Medical Radiation Physics Centre, Radiological and Medical Sciences Research Institute, Ghana Atomic Energy Commission, Accra. Ghana
  • Cyril Schandorf  Radiation Protection Institute, Ghana Atomic Energy Commission, Accra. Ghana

Keywords:

Entrance Surface Dose (ESD), Computed Radiography (CR), Quality control (QC), Diagnostic Reference Level (DRL).

Abstract

This study seeks to investigate the distribution of doses to adult patients undergoing computed radiography using the free-in air method and the factors that affect the doses received by the patients, and hence develop the methodology for optimization of protection of the patient. The entrance surface dose (ESD) was estimated for a total of 619 patients with 243 males (39.3%) and 376 females (60.7%) for two hospitals A and B. The minimum and maximum age of the patients that formed part of this study were 18 and 82 years. The mean ESDs calculated were in the range of 0.29 0.0041 mGy to 6.08 0.55 mGy for chest PA, lumbar spine AP and LAT, cervical spine AP and LAT, skull AP and LAT and abdomen AP examinations. All the results for the various anatomical parts were lower compared with published results and diagnostic reference levels except chest PA examination which the value estimated at hospital B was by a factor of 2.56, 2.20 and 1.65 higher than the mean ESDs values by Inkoom et al, the IAEA and Public Health of UK respectively. An increment by factor of 3.07, 2.87 and 2.15 were found in the estimated mean ESDs for chest PA examination at hospital A compared with the published results respectively. The study therefore has shown that the estimated mean ESD of the hospitals were within the recommended references values except chest PA.

References

  1. Jin, M. G., Im, J. G., Jong H. K., Joon B. S., Tae S. K., Sang J. S., et al. Digital chest radiography with a selenium-based flat-panel detector versus a storage phosphor system: Comparison of soft-copy images. Am J Roentgenology, 2000; 175: 1013-1018.
  2. Williams, M. B., Krupinski, E. A., Strauss, K. J., Breeden, W. K., Rzeszotarski, M. S., Applegate, et al. Digital Radiography Image Quality: Image Acquisition. J ACR. 2007; 4: 371-388.
  3. Inkoom, S., Togobo, J., Emi-reynolds, G., Oddoye, A., Ntiri, T. O., and Gyekye, P. K. Retrospective Patient Dose Analysis of Ghana’s First Direct Digital Radiography System. Health Physics Society. 2012; 103: 133-137.
  4. Kothan, S., and Tungjai, M. An Estimation of X-Radiation Output using Mathematic Model. Am. J App. Sci.2011; 8: 923-926.
  5. Smans, K.; Vandenbroucke, D.; Pauwels, H.; Bosmans, H. Validation an image simulation technique for two computed radiography systems: An application to neonatal imaging. J Med. Phy. 2010; 37: 2092-2100
  6. International Commission on Radiological Protection. Assessing Dose of the Representative Person for the Purpose of the Radiation Protection of the Public. ICRP Publication 101a. Annals of ICRP. 2006; 36.
  7. International Atomic Energy Agency. Dosimetry in Diagnostic Radiology: An International Code of Practice (Technical), Vienna. 2007.
  8. European Commission. European Guidelines on Quality Criteria for Diagnostic Radiographic Images. Luxembourg, 1996.
  9. Institute of Physics and Engineers in Medicine. Recommended Standards for Routine Performance Testing of Diagnostic X-ray Imaging Systems, IPEM Report 91. 2005.
  10. Tamboul, J. Y., Yousef, M., Mokhtar, K., Alfaki, A., and Sulieman, A. Assessment of Entrance Surface Dose for the Patients from Common Radiology Examinations in Sudan. J Life Sci. 2014; 11: 164-168.
  11. Public Health England. Guidance: National Diagnostic Reference Levels (NDRLs). 2016; Retrieved April 30, 2016, from https://www.gov.uk/government/publications/diagnostic-radiology-national-diagnostic-reference-levels-ndrls/national-diagnostic-reference-levels-ndrls#fnref:2
  12. Rehani, M. M. Protection of patients in general radiography. In International Conference 26Y30, Vienna: International Atomic Energy Agency. 2001.
  13. Hess, R., Neitzel, U. Optimizing image quality and dose in digital radiography of paediatric extremities. RoFo-Ortschritte Auf Dem Gebiet Der R, 2012; 184: 643-649.
  14. Reis, C. S., Ndlovu, J., Serrenho, C., Akhtar, I., Haan De, S., Garcia, J., et al. Optimisation of paediatrics computed radiography for full spine curvature measurements using a phantom: a pilot study. Optimax 2014: Radiation Dose and Image Quality Optimisation in Medical Imaging. Retrieved from http://usir.salford.ac.uk/34439/
  15. Doyle, P. Assessment and optimisation of digital radiography systems for clinical use. Assessment and Optimisation of Digital Radiography Systems for Clinical Use. University of Glasgow. 2008: Retrieved from http://theses.gla.ac.uk/873/.
  16. Bath, M. Evaluating Imaging Systems: Practical Applications. Radiation Protection Dosimetry. 2010; 139: 26-36.
  17. International Atomic Energy Agency. Diagnostic Radiology Physics: A Handbook for Teachers and Students, 2014.

Downloads

Published

2017-10-21

Issue

Section

Research Articles

How to Cite

[1]
Daniel Ackom, Edem Sosu, Stephen Inkoom, Cyril Schandorf, " Estimation of Entrance Surface Dose of Adult Patients undergoing Computed Radiography Examinations in Two Hospitals, International Journal of Scientific Research in Science, Engineering and Technology(IJSRSET), Print ISSN : 2395-1990, Online ISSN : 2394-4099, Volume 3, Issue 6, pp.746-752, September-October-2017.